Lina sterilizer

The power of efficiency

Usability outside. Technology inside.

B type sterilizer for best performance.
To prevent the transmission of infections, steam sterilization plays a significant role in the reprocessing of dental instruments. Thereby, efficiency and simplicity are crucial. The new generation of Lina sterilizers meet all these requirements and more: simple operation, ergonomic and functional design, appropriate sterilization cycles and upgradeable functions guarantee high efficiency leading to maximum results.

Sterilization, Hygiene & Maintenance

Sterilization, Hygiene & Maintenance

System Overview

  • Click on the Icon to place the Lina virtually in your reprocessing room and see for yourself how well it fits.
    Drag the mouse in any direction to view the product from all sides.
    Open the website with your mobile phone and experience the Lina in your reprocessing room!


    Lina sterilizer

    B type sterilizer, yet brilliant

    Interrupting the chain of infection.

    B type sterilizer – for peace of mind on reprocessing and infection prevention.

    Benefits at a glance:

    • B type sterilizer with the best cycle times of its category.
    • Ease of use for an unrivalled smooth user experience.
    • Unique upgradeability for customising and modernising the device according to your needs.
    • Advanced traceability for complete documentation.
    • New environmentally friendly packaging The new environmentally friendly packaging is completely reusable and recyclable. Plastic components were reduced to a minimum. For 99% of its weight, the new package is made of cardboard and other fully biodegradable substances.

    Improved components

    Simplified work processes

    Usability outside.
    The Lina sterilizer is characterised by an effortless handling, ergonomic and functional design. Ideal to simplify the sterilization process and thus to ensure an efficient workflow in your practice.

    The risk of infection is actively minimised by sterilization! Because it's all about the safety of your patients, your staff and yourself.


    How may you benefit from the improved usability?

    Ease of use

    Ease of use

    //a.storyblok.com/f/45617/385x289/79f49fe19a/lina-easeofuse_big.jpg
    › The 3.5’’ touch screen display
    enables smooth operation.

    › The simplified user interface and menu structure support practices in all daily activities.
    Optimal sterilization programs

    Optimal sterilization programs

    //a.storyblok.com/f/45617/385x289/39f232e0f0/lina-sterilizationprograms_big.jpg
    › The Eco B sterilization program reprocesses 0.5 kg in 25 minutes.

    › Lina can reprocess up to 6 kg of load
    to meet all requirements of dental clinics.

    › The B Universal 121°C cycle even allows the reprocessing of sensitive items, including porous such as surgical clothing.
    Design and maintenance

    Design and maintenance

    //a.storyblok.com/f/45617/385x289/381a586cb9/lina-designandmaintanance_big.jpg
    › The easily removable cover allows effortless access to the water tanks without tools. Filling is particularly easy with the integrated funnel.

    › Thanks to the high level of reliability, the service interval is set to 4,000 cycles or 5 years.

    › The design impresses with its ergonomic and functional shape and fits in anywhere thanks to modular feet and compact layout.

    Technical concepts

    Realised visions

    Technolgy inside.
    Not only the outside, but also the technical expertise inside makes the difference. The Lina sterilizer is based on a sophisticated technical concept that offers future-oriented possibilities and significantly simplifies your workflow.

    How may you benefit from the advanced technology?


    Use the technological upgrades to master all challenges with your sterilizer – today and tomorrow!
    Enhanced traceability

    Enhanced traceability

    //a.storyblok.com/f/45617/385x289/156647aa97/lina-enhancedtraceability_big.jpg
    › The documentation is automatically
    saved via a high-capacity USB stick.

    › The barcode label printer provides additional support on traceability thanks to the opportunity to link cycle, sterilization package and patient.

    › The cycle report printer ensures adequate paper documentation of the cycle.
    Advanced connectivity

    Advanced connectivity

    //a.storyblok.com/f/45617/385x289/d9a3f4ec1f/lina-connectivity_big.jpg
    › The ioDent® system offers intelligent dental solutions and allows an advanced service support.

    › The W&H Steri App allows controlling and remote monitoring of the sterilizer. Moreover, cycle history back-up saved on smartphone ensures extra-safety.
    Unique upgradeability

    Unique upgradeability

    //a.storyblok.com/f/45617/385x289/64956dc210/lina-upgradeability_big.jpg
    › The Activation Code system is an innovative upgradeability tool. Activation Codes offer a tailored opportunity to enhance additional features that meet your practice needs and comply with future requirements. Effectively, for customising and upgrading your Lina sterilizer.

    Unique upgradeability

    Covering individual needs.

    Needs and requirements may change. Prepare yourself now: with the Activation Code system, additional functions can be activated to meet your changing practice needs and future requirements.

    All in One
    The “All in One” feature provides all benefits at once – and ultimately your sterilizer becomes an unattainable price-performance ratio.

    Six steps to your customized sterilizer:

    1. Purchase Activation Code of your choice from your local approved dental supplier.
    2. After receiving the related login code, go to activation.wh.com
    3. Enter this code along with serial number and model of your sterilizer.
    4. Your individual, secure Activation Code is generated.
    5. Enter this code in the menu of your Lina sterilizer.
    6. The feature is available.


    Lina

    supplied with reversible rack, 3 aluminium trays, drain hose, tray holder, 8 GB USB stick, door opening tool, factory test, 50 Hz

    Lina 17 (RIK-117)
    REF 19951130

    Chamber size 17 l

    Lina 17 AWF
    (RIK-117)

    REF 19951131

    Automatic water filling
    Chamber size 17 l

    Lina 22
    (RIK-122, 220 V)

    REF 19952130

    Chamber size 22 l

    Lina 22 AWF
    (RIK-122, 220 V)

    REF 19952131

    Automatic water filling
    Chamber size 22 l

    Lina 22
    (RIK-122, 110 V)

    REF 19952330

    Chamber size 22 l

    Lina 22 AWF
    (RIK-122, 110 V)

    REF 19952331

    Automatic water filling
    Chamber size 22 l

    Buy

    Lina

    supplied with reversible rack, 3 aluminium trays, drain hose, tray holder, 8 GB USB stick, door opening tool, factory test, 60 Hz

    Lina 17 (RIK-117)
    REF 19951230

    Chamber size 17 l

    Lina 17 AWF
    (RIK-117)

    REF 19951231

    Automatic water filling
    Chamber size 17 l

    Lina 22
    (RIK-122, 220 V)

    REF 19952230

    Chamber size 22 l

    Lina 22 AWF
    (RIK-122, 220 V)

    REF 19952231

    Automatic water filling
    Chamber size 22 l

    Lina 22
    (RIK-122, 110 V)

    REF 19952330

    Chamber size 22 l

    Lina 22 AWF
    (RIK-122, 110 V)

    REF 19952331

    Automatic water filling
    Chamber size 22 l

    Buy

    Download Area

    FAQs

    • What is validation?
      Validation is a testing procedure carried out by an official authority, proving the effectiveness of your sterilizer. Compulsories and frequency of validation are regulated by local/national guidelines. more
    • Do I need to document the results of all periodic tests?
      W&H recommends recording all periodic tests, repairs and modifications on the sterilizer in a logbook including date and signature of the person who carried out the test or other activity. Compulsoriness of recording is regulated by local/national guidelines. more
    • How long should I archive the documentation of regular tests?
      Keep the results according to your country/regional specific laws and guidelines. Those records can be of assistance in case of service as well. more
    • Does my sterilizer need any periodic qualification/validation?
      Validation generally comprises a range of tests defined in internationally recognised standards. These tests begin for every device with the factory tests at W&H. In addition, corresponding documentation is included with your device. Prior to commissioning the device at your premises, the technician performs further checks to ensure the device functions correctly. Then, depending on the national/regional laws and directives, the device must be subjected to a so-called renewed performance qualification at regular intervals. For further information, please contact your dealer. more
    • Whom should I contact to validate and maintain my sterilizer?
      Validation shall be carried out by an official authority. Maintenance tasks shall be carried out by a qualified and authorized technician. more
    • How do I change consumable components on my sterilizer?
      Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available:
      • in the sterilizer, in the maintenance section of the menu; • in the Consumables & Accessories and Video Tutorial section of this web site. more
    • How do I clean the water tanks and what products should I use?
      Water tank cleaning instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available:
      • in the sterilizer, in the maintenance section of the menu;
      • in the Consumables & Accessories and Video Tutorial section of this web site. more
    • Should I pouch wrap my instruments and how long can I store them after sterilization?
      To ensure that your instruments are stored in a sterile state from their successful sterilization until they are used, W&H recommends wrapping them in suitable sterile barrier systems before sterilization. The recommended storage period can be found in the recommendations of the wrapping material manufacturer. For further information, please contact your dealer. more
    • The sterilizer remains switched OFF.
      POSSIBLE CAUSE
      The main switch or network circuit breaker is OFF.
      SOLUTION
      Activate the main switch or network circuit breaker (ON).

      POSSIBLE CAUSE
      No voltage at the socket.
      SOLUTION
      Check the building electrical supply.

      POSSIBLE CAUSE
      The power cord is not connected properly
      SOLUTION
      Check and connect the power cord properly more
    • Water is leaking at the front of the sterilizer.
      POSSIBLE CAUSE
      Leaks through the chamber door seal.
      SOLUTION
      Clean or replace the door seal. Clean the door surface facing the chamber.

      POSSIBLE CAUSE
      Internal leak.
      SOLUTIONS
      Contact an authorized W&H service partner. more
    • At the end of the cycle, there is residual water in the chamber
      POSSIBLE CAUSE
      Sterilizer not properly levelled
      SOLUTION
      Place the sterilizer on a flat, levelled surface.

      POSSIBLE CAUSE
      Overloaded chamber
      SOLUTION
      Comply with the maximum load weight limits for each type of load. Always use the chamber rack for trays and cassettes.

      POSSIBLE CAUSE
      Chamber filter clogged
      SOLUTION
      Remove and clean the chamber filter.

      POSSIBLE CAUSE
      Load incorrectly placed
      SOLUTION
      Follow the recommendations as listed in chapter “Load maintenance and preparation” in the Instructions for Use. more
    • The sterilizer enters into “Standby mode” immediately after being switched on.
      POSSIBLE CAUSE
      The unit was turned off while in standby mode.
      SOLUTION
      Press the “Standby mode” button to exit. more
    • When starting a cycle, the chamber door locks but re-opens immediately. The “Open the door” message appears.
      POSSIBLE CAUSE
      Door seal not properly placed.
      SOLUTION
      Ensure that the door seal is correctly inserted around the whole circumference.

      POSSIBLE CAUSE
      Door jammed by external objects or by the load itself
      SOLUTION
      Remove any objects interfering with the chamber door. Check for interference between the door and the chamber furniture and/or the load. more
    • Stains, spots or colour change on instruments.
      POSSIBLE CAUSE
      Instruments rinsed with tap water and not dried before sterilization.
      SOLUTION
      Ensure that instruments are rinsed with distilled or demineralized water and completely dry before they are placed in the sterilizer.

      POSSIBLE CAUSE
      Use of water of poor quality or water containing chemical substances
      SOLUTION
      Drain both water tanks. Use good quality water (see chapter “Water quality” in the Instructions for Use).

      POSSIBLE CAUSE
      Organic or chemical residues on the instruments
      SOLUTION
      Clean, rinse and dry instruments before placing them in the sterilizer (see chapter “Load maintenance and preparation” in the Instructions for Use).

      POSSIBLE CAUSE
      Dirt in chamber, trays, tray rack.
      SOLUTION
      Clean the chamber and wash the chamber furniture

      POSSIBLE CAUSE
      Contact between instruments of different materials
      SOLUTION
      Ensure that instruments of different materials do not touch each other (aluminum, carbon or stainless steel, etc.); place them on different trays or cassettes or pouch them (see chapter “Load maintenance and preparation” in the Instructions for Use).

      POSSIBLE CAUSE
      Scale deposits on the chamber
      SOLUTION
      Clean the chamber and use good quality water (see chapter “Water quality” in the Instructions for Use). more
    • No cycles are stored in the cycle history menu
      POSSIBLE CAUSE
      An electronic board was replaced when serviced.
      SOLUTION
      The memory of the old board cannot be restored. Consult the cycle history in the USB pen drive. more
    • Warning about saving to the USB (HTML and SCL files)
      POSSIBLE CAUSE
      The USB pen drive is disconnected or full.
      SOLUTION
      Check presence and condition of the USB pen drive. If the problem persist, contact an authorized W&H service partner. more
    • Warning about programmed maintenance.
      POSSIBLE CAUSE
      The maintenance program needs to be carried out.
      SOLUTION
      Contact an authorized W&H service partner. more
    • The cycle report printer does not work.
      POSSIBLE CAUSE
      Printer not properly connected or not powered
      SOLUTION
      Check the data and the power connection to the printer. more
    • The sterilizer is connected to an automated water supply system, there is no clean water in the tank and the automatic water filling does not fill the water.
      POSSIBLE CAUSE
      Water filling system not connected.
      SOLUTION
      Connect the water filling system to the sterilizer. (see Instructions For Use).

      POSSIBLE CAUSE
      When the water filling system attempted to fill the tank, water was temporarily unavailable.
      SOLUTION
      Since the water tank filling is attempted only once in-between cycle execution, this event inhibits water feeding. Switch the sterilizer OFF and then ON again.
      Check the external water supply system.
      Check for water leaks from the sterilizer.

      POSSIBLE CAUSE
      Faulty MIN water level sensor in the clean water tank
      SOLUTION
      Contact an authorized W&H service partner. more
    • The sterilization (PROCESS) phase of a sterilization cycle was longer than expected.
      POSSIBLE CAUSE
      The chamber temperature dropped below the minimum threshold and the software performed a successful recovery
      SOLUTION
      Wait for cycle completion. If the problem occurs frequently, contact an authorized W&H service partner. more
    • Warning about consumable replacement.
      POSSIBLE CAUSE
      A consumable needs to be replaced.
      SOLUTION
      Order the requested consumable (door seal and HEPA/bacteriological filter) and replace it. See the chapter “Maintenance” in the Instructions for Use. more
    • Which is the maximum noise level of Lina?
      The max. noise level of Lina is 65.5 dB. more
    • Which water quality should I use for my Lexa Plus sterilizer?
      The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardised. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Multidem system and OSMO are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use). more
    • Do I need to carry out periodic testing on my W&H Sterilizer?
      All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines. more

    Video Tutorials

    For a seamless and interactive online experience of our multimedia content (e.g. videos, maps), please accept the storage of the necessary cookies (YouTube, MovingImage - detailed information on these cookies can be found in the imprint/data protection) for marketing purposes.

    Accept
  • Lina sterilization cycles

    Sterilization cycle B-Universal 134 B-Prion 134 B-Universal 121 S naked 134 (optional)
    Temperature: 134 °C 134 °C  121 °C 134 °C
    Holding time (minutes): 4' 18' 30'' 20' 30'' 4'
    Total cycle duration
    (minutes) including
    complete drying:
    Automatic from Typical
    load
    Automatic from Typical
    load
    Automatic from Automatic from
    Lina 17
    25’–49’
    45'
    40’–64’
    60‘ 63’–71’ 20’–26’
    Lina 22
    (220 V)
    28’–52’
    48’
    43’–67’
    63’ 65’–74’ 21’–27’
    Lina 22
    (110 V)
    33’–63’
    55’
    47’–77’
    69’ 73’–82’ 28’–35’
    Lina17 loads:
    Lina 22 loads:
    maximum 3.5–5.5(1) kg/typical 2 kg/maximum porous 0.5–1.5(2) kg
    maximum 4–6(3) kg/typical 2 kg/maximum porous 0.5–2(4) kg
    max. 4 kg unwrapped
    max. 4 kg unwrapped
    Lina 17/22 containers: maximum 6 kg
    Test cycles: Helix/Bowie and Dick/Vacuum


    Items supplied with: reversible rack, 3 aluminium trays, drain hose, tray holder, 8 GB USB stick, door opening tool, factory test



    (1) Duration reported in the table refers to 3.5 kg. It is possible to sterilize up to 5.5 kg by drying time extension.
    (2) Duration reported in the table refers to 0.5 kg. It is possible to sterilize up to 1.5 kg by drying time extension.
    (3) Duration reported in the table refers to 4.0 kg. It is possible to sterilize up to 6.0 kg by drying time extension.
    (4) Duration reported in the table refers to 0.5 kg. It is possible to sterilize up to 2.0 kg by drying time extension.

    Lina

    Trade name

    Lina

    Type:

    RIK-117

    RIK-122

    Chamber size:

    17 l

    22 l

    Power supply:

    200–240 V AC; 50 Hz or 60 Hz; 10 A single phase

    200–240 V AC; 50 Hz or 60 Hz; 10 A single phase

    Power consumption:

    2.0–2.4 kW

    220 V: 2.4-2.5 kW
    110 V: 1.2-1.5 kW

    Overall dimension (w x h x d):

    465 × 452 × 646 mm

    Weight (empty):

    40 kg

    42 kg

    Main/Used water tanks:

    4.8/4.8 l

    Working range:

    7 to 10 cycles

    Usable space in chamber (w × h × d)¹:

    195 × 195 × 312 mm

    195 × 195 × 400 mm

    Connection types:

    1 USB port in the front (additional 1 in the rear as option), automatic water filling kit (optional)

    (1) Usable space is referred to standard rack configuration.

    Lina sterilizers were designed, certified and validated with the most stringent directives and standards:

    N°2017/745

    Medical Device Regulation

    IEC 61326-1

    Electromagnetic compatibility

    2014/68/EU

    Pressure equipment Directive

    IEC 61010-1

    Safety requirements sterilizer

    EN 13060

    Small steam sterilizer

    IEC 61010-2-040

    Specific requirements for steam sterilizer

    IEC 61770

    Electric appliances connected to the water mains

    2012/19/EU

    Waste Electrical and Electronic Equipment


    The sterilizer can be validated according to EN ISO 17665-1