Interrupting the
chain of infectionfor peace of mind on reprocessing and infection prevention.
The Lexa Plus is built to offer more of all that counts in everyday dental infection control. It fits perfectly into each individual practice environment and workflow.
Environmentally friendly packaging
The new environmentally friendly packaging is completely reusable and recyclable. Plastic components were reduced to a minimum. For 99% of its weight, the new package is made of cardboard and other fully biodegradable substances.Innovative, reliable and fast
- Eco Dry + technology, fast cycles and energy saving
- High capacity, 18 lbs. maximum load and optimized rack holds 5 trays
- Enhanced traceability
- Convenient, automatic water filling saves time
- Class B, pre-vacuum sterilizer
- Versatile cycle program options for all types of loads
- Closed-door drying with HEPA filter
-
Single-use distilled water management, ensuring fresh water for every cycle
Reprocessing in less than 22 minutes
Thanks to Eco Dry +, Lexa Plus cycles are very efficient: only 20 minutes for unwrapped loads and
only 8 minutes more for wrapped loads.Reprocessing in less than 22 minutes!
Wipe disinfection in 1 minute
Disinfect your instruments by wiping them down before cleaning and maintenance.
10 seconds maintenance
The Assistina Twin reduces the maintenance process to a record time of 10 seconds.
Sterilization in 20 minutes
With the Lexa Plus fast cycle, instruments can be removed from the sterilizer ready for use after only
20 minutes.
Higher efficiency due to larger capacity …
Lexa Plus comes with 5 large trays along with a unique rack providing larger loading surface and more loading options.... and less cycle time.
Lexa Plus offers versatile program options for all types of loads in order to reduce overall cycle duration.
Protection at a new level
with EliTraceAdvanced traceability – Set a new standard with EliTrace
Intuitive navigation
The intelligent menu structure of the new Lexa Plus sterilizer ensures extraordinarily high operating comfort.Intuitive navigation
You can navigate to the sterilization programs quickly and easily using the large 4.3” color display.
Water management system
An automatic water filling valve allows you to connect a demineralized system for optimal efficiency.
Manual filling and draining are no longer necessary.Easy filling
Thanks to a frontal integrated funnel, manual filling is possible and easy.
Effortless tank access
The tank cover can be easily removed without tools to clean the tank.
Available Accessories
Compatibility:
The application is compatible with all sterilizers currently available.
Available languages: EN, IT, FR, DE, ES, ZH-CN, JP
Main features:
Remote control of your sterilizer: the user can remotely command up to 4 sterilizers.
Advanced traceability: EliTrace feature is now fully supported and the App permits to create instrument database directly from your smart device. In addition to this function, the management and tracking of dental instruments is integrated.
Reports creation: the App allows making intervention reports in order to collect data regarding specific treatments.
User management: creation and handling of sterilizer users directly from your smart device.
Backup functionality: all the cycle reports are automatically saved to assure more protection of data.
Lexa Plus sterilizer
supplied with rack, 5 aluminum trays, drain hose, tray holder, 8 GB USB stick, door opening tool, factory test
Lexa Plus
RIS-311
220 V
REF 199331007.4 gal (28 liters)
200–240 V AC; 50/60 Hz 10 A Single PhaseDownload Area
FAQs
-
What is validation?
Validation is a testing procedure carried out by an official authority, proving the effectiveness of your sterilizer. Compulsories and frequency of validation are regulated by local/national guidelines. more
-
Do I need to document the results of all periodic tests?
W&H recommends recording all periodic tests, repairs and modifications on the sterilizer in a logbook including date and signature of the person who carried out the test or other activity. Compulsoriness of recording is regulated by local/national guidelines. more
-
How long should I archive the documentation of regular tests?
Keep the results according to your country/regional specific laws and guidelines. Those records can be of assistance in case of service as well. more
-
Does my sterilizer need any periodic qualification/validation?
Validation generally comprises a range of tests defined in internationally recognised standards. These tests begin for every device with the factory tests at W&H. In addition, corresponding documentation is included with your device. Prior to commissioning the device at your premises, the technician performs further checks to ensure the device functions correctly. Then, depending on the national/regional laws and directives, the device must be subjected to a so-called renewed performance qualification at regular intervals. For further information, please contact your dealer. more
-
Whom should I contact to validate and maintain my sterilizer?
Validation shall be carried out by an official authority. Maintenance tasks shall be carried out by a qualified and authorized technician. more
-
How do I change consumable components on my sterilizer?
Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available:
• in the sterilizer, in the maintenance section of the menu; • in the Consumables & Accessories and Video Tutorial section of this web site. more -
How do I clean the water tanks and what products should I use?
Water tank cleaning instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available:
• in the sterilizer, in the maintenance section of the menu;
• in the Consumables & Accessories and Video Tutorial section of this web site. more -
Should I pouch wrap my instruments and how long can I store them after sterilization?
To ensure that your instruments are stored in a sterile state from their successful sterilization until they are used, W&H recommends wrapping them in suitable sterile barrier systems before sterilization. The recommended storage period can be found in the recommendations of the wrapping material manufacturer. For further information, please contact your dealer. more
-
The sterilizer remains switched OFF.
POSSIBLE CAUSE
The main switch or network circuit breaker is OFF.
SOLUTION
Activate the main switch or network circuit breaker (ON).
POSSIBLE CAUSE
No voltage at the socket.
SOLUTION
Check the building electrical supply.
POSSIBLE CAUSE
The power cord is not connected properly
SOLUTION
Check and connect the power cord properly more -
Water is leaking at the front of the sterilizer.
POSSIBLE CAUSE
Leaks through the chamber door seal.
SOLUTION
Clean or replace the door seal. Clean the door surface facing the chamber.
POSSIBLE CAUSE
Internal leak.
SOLUTIONS
Contact an authorized W&H service partner. more -
At the end of the cycle, there is residual water in the chamber
POSSIBLE CAUSE
Sterilizer not properly levelled
SOLUTION
Place the sterilizer on a flat, levelled surface.
POSSIBLE CAUSE
Overloaded chamber
SOLUTION
Comply with the maximum load weight limits for each type of load. Always use the chamber rack for trays and cassettes.
POSSIBLE CAUSE
Chamber filter clogged
SOLUTION
Remove and clean the chamber filter.
POSSIBLE CAUSE
Load incorrectly placed
SOLUTION
Follow the recommendations as listed in chapter “Load maintenance and preparation” in the Instructions for Use. more -
The sterilizer enters into “Standby mode” immediately after being switched on.
POSSIBLE CAUSE
The unit was turned off while in standby mode.
SOLUTION
Press the “Standby mode” button to exit. more -
When starting a cycle, the chamber door locks but re-opens immediately. The “Open the door” message appears.
POSSIBLE CAUSE
Door seal not properly placed.
SOLUTION
Ensure that the door seal is correctly inserted around the whole circumference.
POSSIBLE CAUSE
Door jammed by external objects or by the load itself
SOLUTION
Remove any objects interfering with the chamber door. Check for interference between the door and the chamber furniture and/or the load. more -
Stains, spots or colour change on instruments.
POSSIBLE CAUSE
Instruments rinsed with tap water and not dried before sterilization.
SOLUTION
Ensure that instruments are rinsed with distilled or demineralized water and completely dry before they are placed in the sterilizer.
POSSIBLE CAUSE
Use of water of poor quality or water containing chemical substances
SOLUTION
Drain both water tanks. Use good quality water (see chapter “Water quality” in the Instructions for Use).
POSSIBLE CAUSE
Organic or chemical residues on the instruments
SOLUTION
Clean, rinse and dry instruments before placing them in the sterilizer (see chapter “Load maintenance and preparation” in the Instructions for Use).
POSSIBLE CAUSE
Dirt in chamber, trays, tray rack.
SOLUTION
Clean the chamber and wash the chamber furniture
POSSIBLE CAUSE
Contact between instruments of different materials
SOLUTION
Ensure that instruments of different materials do not touch each other (aluminum, carbon or stainless steel, etc.); place them on different trays or cassettes or pouch them (see chapter “Load maintenance and preparation” in the Instructions for Use).
POSSIBLE CAUSE
Scale deposits on the chamber
SOLUTION
Clean the chamber and use good quality water (see chapter “Water quality” in the Instructions for Use). more -
No cycles are stored in the cycle history menu
POSSIBLE CAUSE
An electronic board was replaced when serviced.
SOLUTION
The memory of the old board cannot be restored. Consult the cycle history in the USB pen drive. more -
Warning about saving to the USB (HTML and SCL files)
POSSIBLE CAUSE
The USB pen drive is disconnected or full.
SOLUTION
Check presence and condition of the USB pen drive. If the problem persist, contact an authorized W&H service partner. more -
Warning about programmed maintenance.
POSSIBLE CAUSE
The maintenance program needs to be carried out.
SOLUTION
Contact an authorized W&H service partner. more -
The cycle report printer does not work.
POSSIBLE CAUSE
Printer not properly connected or not powered
SOLUTION
Check the data and the power connection to the printer. more -
The sterilizer is connected to an automated water supply system, there is no clean water in the tank and the automatic water filling does not fill the water.
POSSIBLE CAUSE
Water filling system not connected.
SOLUTION
Connect the water filling system to the sterilizer. (see Instructions For Use).
POSSIBLE CAUSE
When the water filling system attempted to fill the tank, water was temporarily unavailable.
SOLUTION
Since the water tank filling is attempted only once in-between cycle execution, this event inhibits water feeding. Switch the sterilizer OFF and then ON again.
Check the external water supply system.
Check for water leaks from the sterilizer.
POSSIBLE CAUSE
Faulty MIN water level sensor in the clean water tank
SOLUTION
Contact an authorized W&H service partner. more -
The sterilization (PROCESS) phase of a sterilization cycle was longer than expected.
POSSIBLE CAUSE
The chamber temperature dropped below the minimum threshold and the software performed a successful recovery
SOLUTION
Wait for cycle completion. If the problem occurs frequently, contact an authorized W&H service partner. more -
Warning about consumable replacement.
POSSIBLE CAUSE
A consumable needs to be replaced.
SOLUTION
Order the requested consumable (door seal and HEPA/bacteriological filter) and replace it. See the chapter “Maintenance” in the Instructions for Use. more -
What is the maximum noise level of the Lexa Plus sterilizer?
The max. noise level of the Lexa Plus is 70 dB. more
-
Which water quality should I use for my Lexa Plus sterilizer?
The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardised. Consequently, only water with conductivity of less than 10 µS/cm in North America and 15 µS/cm for the rest of the world should be used for your W&H sterilizer. The Multidem system and OSMO are especially designed to prepare suitable water (see chapter “Water quality” in the Instructions for Use). more
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines. more
Instructions for use
Find all languages of the Update Hygiene and Maintenance (EN ISO 17664:2017) in the Download Centre.
Video Tutorials
-
Lexa Plus Sterilization Programs
Sterilization cycles: Wrapped Low Temperature Unwrapped Temperature: 270°F (132°C) 250°F (121°C) 270°F (132°C) Sterilization time: 4 min 30 min 4 min Total cycle duration
(minutes) including
complete drying time:Automatic fromTypical load Automatic fromTypical load Automatic from Typical load 28–62 min39 min 68–107 min85 min 20–40 min 24 min Max. load: 14 lbs (6.3 kg)
Pouched/unwrapped items on trays18 lbs (8.1 kg)
Unwrapped instruments and dental handpieces on trays
16.5 lbs (7.5 kg)
Items in wrapped/unwrapped
cassettes or in trays and wrapped/unwrapped cassettes4.4 lbs (2.0 kg)
Pouched/unwrapped Textile load on traysSterilizer supplied with rack, 5 aluminum trays, drain hose, tray holder, 8 GB USB stick, door opening tool, factory test Lexa Plus
Trade name:
Type:Lexa Plus
RIS-311
Chamber size: 7.4 gal (28 liters) Power supply: 200–240 V AC; 50/60 Hz; 10 A single phase Power consumption: 2.0–2.4 kW Overall dimension
(width × height × depth):19.3“ × 17.9“ × 24.4“ (492 × 455 × 620 mm) Weight (empty): 119 lbs (54 kg) Main/used water tanks: 1.27 gal/1.35 gal (4.8 liters/5.11 liters) Working range: 5 to 15 cycles Usable space in chamber
(width × height × depth)¹:9“ × 9“ × 17.3“ (230 × 230 × 439.4 mm) Connection types: 5 USB ports, 1 Ethernet (LAN) port, automatic water filling kit (1) Usable space is referred to standard rack configuration.
Lexa Plus sterilizers were designed, certified and validated with the most stringent directives and standards:
EN 13060
Small steam sterilizers
IEC 61326-1
Electromagnetic compatibility
ASME
Boiler and pressure vessel code
IEC 61010-1
Safety requirements
ANSI/AAMI ST55
Table-top steam sterilizers
IEC 61010-2-040
Specific requirements for steam sterilizer
N°2017/745
Medical Device Regulation
IEC 61770
Electric appliances connected to the water mains (avoidance of backsiphonage)
2014/68/EU
Pressure Equipment Directive
2012/19/EU
Waste Electrical and Electronic Equipment (WEEE)
CSA-C22.2 No 61010-1
Safety requirements for electrical
equipment for measurement, control and laboratory use, general requirementsUL 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use, general requirements
› The sterilizer can be validated according to EN ISO 17665-1