Lara Sterilizers
Ready for the futureIncredible outside
You don’t need to be a lover of design to appreciate the Lara sterilizers. Their smooth surfaces are comfortable to the touch, the instruments are easy to load and the flat colour screen displays allow you to navigate through the intelligent menu structure quickly and with ease.
Incredible inside
Discover the W&H Lara sterilizers down to the smallest detail. So many components, perfectly arranged to offer you complete satisfaction by simplifying the sterilization process while providing you with comprehensive documentation. Unique upgradeability allows you to customise your Lara sterilizer with your personal needs and to comply with regulatory requirements.Incredible because
- Fastest B cycle in its segment
- Unique upgradeability thanks to Activation Codes
- Colour touch display for easy navigation
- Automatic cycle record via high capacity USB stick
- Optional high end label and cycle report printer
- Ergonomics and functional design
- Ideal dimensions to integrate and operate
-
New environmentally friendly packaging
The new environmentally friendly packaging is completely reusable and recyclable. Plastic components were reduced to a minimum. For 99% of its weight, the new package is made of cardboard and other fully biodegradable substances.
Six steps to your customized sterilizer:
- Purchase Activation Code of your choice from your local approved dental supplier.
- After receiving the related login code go to activation.wh.com
- You specify this code along with the serial number and model of your sterilizer.
- Your individual, secure Activation Code is generated.
- Enter this code in the menu of your Lara sterilizer.
- The feature is available.
Reprocessing in less than 23 minutes!
Lara sterilizer
supplied with reversible rack, 3 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test
Lara 17
REF 19941122Chamber size 17 l
Lara 22
REF 19942122Chamber size 22 l
Download Area
FAQ
-
The sterilizer remains switched OFF.
POSSIBLE CAUSE
The main switch or network circuit breaker is OFF.
SOLUTION
Activate the main switch or network circuit breaker (ON).
POSSIBLE CAUSE
No voltage at the socket.
SOLUTION
Check the building electrical supply.
POSSIBLE CAUSE
The power cord is not connected properly.
SOLUTION
Check and connect the power cord properly. more -
Should I pouch/wrap my instruments and how long can I store them after sterilization?
To ensure that your instruments are stored in a sterile state from their successful sterilization until they are used, W&H recommends wrapping them in suitable sterile barrier systems before sterilization. The recommended storage period can be found in the recommendations of the wrapping material manufacturer. For further information, please contact your dealer. more
-
Which water quality should I use for my sterilizer?
The water quality for sterilizers is internationally specified by standards and directives. The reasons for this are that your sterilizer can be seriously damaged by it or the sterilization can be jeopardized. Consequently, only water with conductivity of less than 15 µS/cm should be used for your W&H sterilizer. The Water treatment devices is especially suited to preparing suitable water (see chapter “Water quality” in the Instructions for Use). more
-
Whom should I contact to validate and maintain my sterilizer?
Validation shall be carried out by an official authority. Maintenance tasks shall be carried out by a qualified and authorised technician. more
-
Does my sterilizer need any periodic qualification/validation?
Validation generally comprises a range of tests defined in internationally recognised standards. These tests begin for every device with the factory tests at W&H. In addition, corresponding documentation is included with your device. Prior to commissioning the device at your premises, the technician performs further checks to ensure the device functions correctly. Then, depending on the national/regional laws and directives, the device must be subjected to a so-called renewed performance qualification at regular intervals. For further information, please contact your dealer. more
-
How long should I archive the documentation of regular tests?
Keep the results according to your country/regional specific laws and guidelines. Those records can be of assistance in case of service as well. more
-
Do I need to document the results of all periodic tests?
W&H recommends recording all periodic tests, repairs and modifications on the sterilizer in a logbook including date and signature of the person who carried out the test or other activity. Compulsories of recording is regulated by local/national guidelines. more
-
At the end of the cycle, there is residual water in the chamber.
POSSIBLE CAUSE
Sterilizer not properly levelled.
SOLUTION
Place the sterilizator on a flat, level surface.
POSSIBLE CAUSE
Overloaded chamber.
SOLUTION
Comply with the maximum load weight limits for each type of load. Always use the chamber rack for trays and cassettes.
POSSIBLE CAUSE
Chamber filter clogged.
SOLUTION
Remove and clean the chamber filter.
POSSIBLE CAUSE
Chamber filter cap mispositioned.
SOLUTION
Mount the chamber filter cap properly (see chapter “Monthly or 50-cycle maintenance - Cleaning the chamber filter” in the Instructions for Use).
POSSIBLE CAUSE
Load incorrectly placed.
SOLUTION
Follow the recommendations as listed in chapter “Load maintenance and preparation” in the Instructions for Use. more -
Water is leaking at the front of the sterilizer.
POSSIBLE CAUSE
Leaks through the chamber door seal.
SOLUTION
Clean or replace the door seal. Clean the door surface facing the chamber.
POSSIBLE CAUSE
Internal leak.
SOLUTION
Contact an authorized W&H service partner. more -
The cycle report printer does not work.
POSSIBLE CAUSE
Printer not properly connected or not powered.
SOLUTION
Check the data and power connection to the printer. more -
The sterilizer is connected to an automated water supply system, there is no clean water in the tank and the automatic water filling does not fill the water.
POSSIBLE CAUSE
Water filling system not connected.
SOLUTION
Connect the water filling system to the sterilizer. (see ANNEX 7 For water quality requirements see chapter “Water quality” in the Instructions For Use).
POSSIBLE CAUSE
When the water filling system attempted to fill the tank, water was temporarily unavailable.
SOLUTION
Since the water tank filling is attempted only once in-between cycle execution, this event inhibits water feeding. Switch the sterilizer OFF and then ON again.
Check the external water supply system.
Check for water leaks from the sterilizer.
POSSIBLE CAUSE
Faulty MIN water level sensor in the clean water tank.
SOLUTION
Contact an authorized W&H service partner. more -
Warning about programmed maintenance.
POSSIBLE CAUSE
The maintenance program needs to be carried out.
SOLUTION
Contact an authorized W&H service partner. more -
The sterilization (PROCESS) phase of a sterilization cycle was longer than expected.
POSSIBLE CAUSE
The chamber temperature dropped below the minimum threshold and the software performed a successful recovery.
SOLUTION
Wait for cycle completion. If the problem occurs frequently, contact an authorised W&H service partner. more -
The sterilizer enters into “Standby mode” immediately after being switched on.
POSSIBLE CAUSE
The unit was turned off while in standby mode.
SOLUTION
Press the standby mode button to exit. more -
When starting a cycle, the chamber door locks but re-opens immediately. The “Open the door” message appears.
POSSIBLE CAUSE
Door seal not properly placed.
SOLUTION
Ensure that the door seal is correctly inserted around the whole circumference.
POSSIBLE CAUSE
Door jammed by external objects or by the load itself.
SOLUTION
Remove any objects interfering with the chamber door. Check for interference between the door and the chamber furniture and/or the load. more -
No cycles are stored in the cycle history menu.
POSSIBLE CAUSE
An electronic board was replaced when serviced.
SOLUTION
The memory of the old board can not be restored. Consult the cycle history in the USB pen drive. more -
Stains or spots on instruments.
POSSIBLE CAUSE
Instruments rinsed with tap water and not dried before sterilization.
SOLUTION
Ensure that instruments are rinsed with distilled or demineralized water and completely dry before they are placed in the sterilizer.
POSSIBLE CAUSE
Use of water of poor quality or water containing chemical substances.
SOLUTION
Drain both water tanks. Use good quality water (see chapter “Water quality” in the Instructions for Use).
POSSIBLE CAUSE
Organic or chemical residues on the instruments
SOLUTION
Clean, rinse and dry instruments before placing them in the sterilizer (see chapter “Load maintenance and preparation” in the Instructions for Use.
POSSIBLE CAUSE
Dirt in chamber, trays, tray rack.
SOLUTION
Clean the chamber and wash the chamber furniture.
POSSIBLE CAUSE
Contact between instruments of different materials.
SOLUTION
Ensure that instruments of different materials do not touch each other (aluminium, carbon or stainless steel, etc.); place them on different trays or cassettes or pouch them (see chapter “Load maintenance and preparation” in the Instructions for Use).
POSSIBLE CAUSE
Scale deposits on the chamber.
SOLUTION
Clean the chamber and use good quality water (see chapter “Water quality” in the Instructions for Use). more -
How do I clean the water tanks and what products should I use?
Water tank cleaning instructions are provided in the Instructions for Use, chapter “Maintenance”. Information and videos relating to the cleaning of the water tanks and cleaning products can also be found on the Video Tutorial section of this web site. more
-
What is the maximum noise level of the Lara sterilizer?
The max. noise level of Lara is 66.9 dB. more
-
Do I need to carry out periodic testing on my W&H Sterilizer?
All W&H sterilizers features test cycles (B&D/Helix test, Vacuum test). Compulsory and frequency of testing are regulated by local/national guidelines. more
-
Warning about consumable replacement.
POSSIBLE CAUSE
A consumable needs to be replaced.
SOLUTION
Order the requested consumable (door seal, dust filter, bacteriological filter) and replace it. See the chapter “Maintenance” in the Instructions for Use. more -
How do I change consumable components on my sterilizer?
Consumable replacement instructions are provided in the Instructions for Use, chapter “Maintenance”. In addition, a video description is available: • in the sterilizer, in the maintenance section of the menu (only for Lisa); • in the Consumables & Accessories and Video Tutorial section of this web site. more
-
Warning about saving to the USB (HTML and SCL files)
POSSIBLE CAUSE
The USB mass storage is disconnected or full.
SOLUTION
Check presence and condition of the USB pen drive. If the problem persist, contact an authorized W&H service partner. more -
What is validation?
Validation is a testing procedure carried out by an official authority, proving the effectiveness of your sterilizer. Compulsories and frequency of validation are regulated by local/national guidelines. more
Instructions for use
Find all languages of the Update Hygiene and Maintenance (EN ISO 17664:2017) in the Download Centre.
Videos & Tutorials
-
Lara sterilization cycles
Sterilization cycle B-Universal 134 B-Prion 134 B-Universal 121 S FAST 134 (optional) Temperature: 134 °C 134 °C 121 °C 134 °C Holding time (minutes): 4 18.5 20.5 3.5 Total cycle duration
(minutes) including
complete drying:Lara 17 Automatic from25 to 5137 Automatic from40 to 6552 41 to 73 20 to 22 Lara 22 Automatic from28 to 5539 Automatic from40 to 6854 42 to 76 20 to 22 Lara 17 loads:Lara 22 loads:maximum solid 4.5 kg/typical 2 kg/maximum porous 1.5 kgmaximum solid 5.5 kg/typical 2 kg/maximum porous 2 kgmax. unwrapped 2 kgmax. unwrapped 2 kgLara 17/22 containers load: maximum 9 kg maximum 9 kg Test cycles: Helix/Bowie Dick/Vacuum test Items supplied with: reversible rack, 3 aluminium trays, drain tube, tray holder, 8 GB USB drive, door opening tool, factory test
Lara
Trade name: Lara Type: RIL-117 RIL-122 Chamber size: 17 l 22 l Power supply: 200–240 V AC; 50/60 Hz; 10 A
Power consumption: 2.0–2.4 kW Overall dimension
(width x height x depth):465 x 452 x 646 mm Weight: 42.5 kg 44 kg Main / Used water tanks: 4.8/4.8 l Working range: 7 to 15 cycles Usable space in chamber
(width x height x depth)¹:195 x 195 x 312 mm 195 x 195 x 400 mm Connection types: 2 USB ports (5 as option), integrated automatic water filling valve (1) Usable space is referred to standard rack configuration.
Lara sterilizers were designed, certified and validated with the most stringent directives and standards:
N°2017/745
Medical Device Regulation
IEC 61326-1
Electromagnetic compatibility
2014/68/EU
Pressure equipment Directive
IEC 61010-1
Safety requirements sterilizer
EN 13060
Small steam sterilizer
IEC 61010-2-040
Specific requirements for steam sterilizer
IEC 61770
Electric appliances connected to the water mains
2012/19/EU
Waste Electrical and Electronic Equipment
The sterilizer can be validated according to EN ISO 17665-1